Specifications |
Pumping Mechanism: |
Piston Mechanism |
Flow Rate: |
1 to 500 ml/hr in 1ml increments |
Total Infused Volume: |
1 to 10,000 ml. |
Total Time Setting |
100 hours |
Accuracy: |
� 5%. |
KTO rate: |
1 to 5 ml/hr. (to be set) |
Air Sensor: |
Ultrasonic, adjustable air bubble size |
Maximum Pressure: |
15 psi Adjustable (high/normal/low) |
Power Supply: |
110-240 VAC, 50/60 Hz. 0.12A |
Battery: |
Li-Ion 7.2V, 1450 mA. (Rechargeable) |
Battery Operation at 125 ml/hr |
15 hours |
Battery Charging:
|
Automatic when clicked into the Charger connected to an AC power source (4 hours) |
Language: |
English or German |
Alarms: |
Mechanical Error
Low & End Battery
Pump Unattended
Air in line
Down Occlusion
Door Open
Program End
Locking state |
Acoustic buzzer level: |
Low to high � adjustable. |
Dimensions: |
112 x 89 x 32mm. (L x w x h). |
Symbol |
Attention (Consult accompanying documents)
|
Surrounding
|
Min. +150C Max. 350C. Humidity 85%. |
|
|
|
|
Classification
|
Type CF Equipment
(degree of protection against electrical shock)
Class IIa device (according to 93/42/EEC Medical devices directive)
|
Housing: |
ABS (fire retardant)
|
Weight
|
280 gr. (without battery)
|
Electrical Safety |
Complies with:
IEC 601-1 (Medical Electrical Equipment Safety), IEC601-2-24 (Infusion pumps and controllers),
IEC 601-1-4 (Programmable Electrical Medical System), UL 2601- 1 and CAN/CSA C22.2 No 601.1.
|
Standards |
Manufacture in accordance to ISO 9001 and EN 46001
CE marked (In accordance with the Medical Devices Directive 93/42/EEC)
|
Special Information: |
Service reminder after 5000 hours.
Call off the last 256 events.
Nurse call with wire or wireless � LPS (optional) |