Technical Specification
BodyGuard 545 Color Vision™
Piston Mechanism

Linear Syringe Driver Mechanism
Infusion Modes
Basal Only

Basal & Bolus

Bolus Only

Auto Bolus

PiAB

Drug Library
Number of Drug Protocols

26 Protocols
Syringe Size*
Syringe Brands
Concentration Units
mg/mcg
Flow Rate Accuracy
±5 %
Basal Rate Range
0.1-30ml/h; 0.1 ml/h Steps
Adjustable Maximum Rate

Maximum Bolus Rate
1200 ml/h
Volume to be Infused
0.1-99.9 ml; 0.1 ml Steps
100-9999 ml; 1 ml Steps
Max Syringe Travel
KVO Rate
0(OFF) - 5 ml/h
Resume on Restart

Ultrasonic Air Sensor
Single Bubble
OFF, 0.1 – 2 ml
Accumulated Air-in-Line

Occlusion Pressure
350– 1100 mmHg; 10mmHg Steps
PCA
PCEA
Regional
Pressure Sensor Mechanism

Actual Pressure Display & Graph

Post-Occlusion Bolus Reduction

Pump Alarms/Alerts
Air/Up Occlusion

Down Occlusion

End Program

Near End
End Syringe Travel
Low Battery

End Battery

Door Open

Pump Unattended

Lock Mode

Missing Key

Syringe Displaced
Check Program

Check Syringe Placement
Display
Size (W/H)
3.7x4.9 cm
Power Supply
100 – 240V AC, 50/60 Hz (10W / 0.3A)

Resolution (W/H)
240x320 pixels
Battery
Rechargeable Li-Polymer 7.4V 1800 mAh.
Life at 125ml/h:
Recharge Time (Empty to Full):
Up to 17 hrs
6hrs

Disposable Alkaline 9V (x2).
Life at 125ml/h:
Up to 12 hrs

Extended-life Rechargeable Li-Polymer 7.4 V
3600 mAh. Life at 125ml/h:
Recharge Time (Empty to Full)

Up to 34 hrs
10hrs
Connectivity (Optional)
RS232 Serial Port

802.11 WiFi

BodyComm™ Software (Optional)

Remote Monitoring System (Optional)

EMR Interface

Classification
Type CF, Class II
Moisture Protection
IPX3
Dimensions; pump only (H/W/D)
112 x 89 x 40 mm
Weight (with standard battery)
390g
* Max volume infused for single infusion varies between syringe brands.
** Specification may vary depending on software version.
*** Manufacturer reserves the right to change the specifications without prior notification.