Technical Specification

BodyGuard 323MR™ 

Piston Mechanism

Infusion Modes

     Continuous

     Dose

          Concentration Units

mg/ml, U/ml

          Dose Units

mg/kg/min or hr

mcg/kg/min or hr

U/kg/min or hr

mU/kg/min or hr

     Non-Weight Dose

          Concentration Unit

mg/ml, U/ml

          Dose Units

mg/min or hr

mcg/min or hr

U/min or hr

mU/min or hr

Delay Before Start 

Time/Volume Mode

Flow Rate Accuracy

± 5 %

Flow Rate Range

0.1-99.9 ml/h; 0.1ml/h Steps

100-1200 ml/h; 1ml/h Step

Adjustable Maximum Rate

Maximum Bolus Rate

1200 ml/h

Programmable Infusion Time

Up to 100 hrs

Volume to be Infused

0.1-99.9 ml; 0.1 ml Steps

100-9999 ml; 1 ml Steps

 Maximum Bolus Volume 

100 ml

KVO Rate

0 (OFF) - 5 ml/h

Resume on Restart

Ultrasonic Air Sensor

     Single Bubble

0.1-1 ml

     Accumulated Air-in-Line

Occlusion Pressure

5, 7.5 or 10 psi

Pressure Sensor Mechanism

Post-Occlusion Bolus Reduction

Pump Alarms/Alerts

     Air-in-Line

     Down Occlusion

     Low Battery

     End Program

     End Battery

     Pump Unattended

     Door Open

     Lock Mode

     Missing Key

Display

     Size (W/H)

6x2 cm

     Resolution (W/H)

122x32 pixels

Power Supply

100 – 240V AC, 47-63 Hz (5.0A)

Battery

Rechargeable Li-Polymer 7.4V 1800 mAh.

Life at 125ml/h:

Up to 15hrs

Connectivity (Optional)
 

      RS232 Serial Port 

      802.11 WiFi 

 BodyComm™ Software (Optional) 

 Remote Monitoring System (Optional) 

 EMR Interface 

Protection Against Electrical Shock

     Pump/Bracket 

Type CF (internally powered

     External Power Supply

Type CF, Class I

 Moisture Protection 

IPX1

Dimensions; pump only (H/W/D)

112 x 89 x 40 mm

 Weight (with standard battery) 

390g

* Specification may vary depending on software version.
** Manufacturer reserves the right to change the specifications without prior notification.