Technical Specification

BodyGuard 575™ 

Piston Mechanism

Linear Syringe Driver Mechanism

Infusion Modes

     Basal Only

     Basal & Bolus

     Bolus Only

     Auto Bolus

Drug Library

Display

     Size (W/H)

6x2 cm

     Resolution (W/H)

122x32 pixels

Flow Rate Accuracy

±5 %

Syringe Size*

Syringe Brands

Flow Rate Range

0.1 – 100 ml/h

Volume to be Infused

0.1 - 9999 ml

Max Syringe Travel

KVO Rate

0(OFF) - 5 ml/h

Ultrasonic Air Sensor

Air bubble size 0.1 - 1 ml

Occlusion Pressure

5, 7.5, 10 psi

     PCA

     PCEA

     Regional

Post-Occlusion Bolus Reduction

Pump Alarms/Alerts

     Air-in-Line

     Occlusion (up/down stream)

     End Program

     Near End

     End Syringe Travel

     Low Battery

     End Battery

     Door Open

     Pump Unattended

     Lock Mode

     Missing Key

     Syringe Displaced

     Check Program

     Check Syringe Placement

Power Supply 100 – 240V AC,
50/60 Hz (10W / 0.3A)

Battery

     Rechargeable Li-Polymer 7.4V 1800
     mAh. Life at 125ml/h:

Up to 15 hrs

     Extended-life Rechargeable Li-
     Polymer 7.4 V 3600 mAh.
     Life at 125ml/h:

Up to 30 hrs

     Disposable Alkaline 9V (x2).
     Life at 125ml/h:

Up to 10 hrs

Connectivity (Optional)

     RS232 Serial Port

     802.11 WiFi

BodyComm™ Software (Optional)

Remote Monitoring System (Optional)

EMR Interface

Classification

Type CF, Class II

Moisture Protection

IPX3

Dimensions; pump only (H/W/D)

112 x 89 x 40 mm

Weight (with standard battery)

390 g

* Max volume infused for single infusion varies between syringe brands.
** Specification may vary depending on software version.
*** Manufacturer reserves the right to change the specifications without prior notification.